Description
Overview
The Siemens SOMATOM Definition DS revolutionized diagnostic imaging by introducing the world’s first dual-source CT architecture. By placing two complete X-ray sources and detector systems on a single rotating gantry, the system cuts the required angular data collection time in half. This breakthrough overcomes the physical limitations of single-tube rotation speeds, making motion-free cardiac imaging routine.
Designed for high-acuity hospitals and specialized cardiology centers, the Definition DS handles complex diagnostic scenarios—such as high heart rates, severe arrhythmias, and uncooperative trauma patients—without requiring pre-medication like beta-blockers. Furthermore, its native Dual-Energy capability allows clinicians to scan at two different energy spectra simultaneously, revealing hidden tissue characteristics, bone removal maps, and contrast agent concentrations without a dose penalty.
Key Features & Advanced Technologies
-
Dual-Source Geometry: Utilizes two pairs of X-ray tubes and UFC (Ultra Fast Ceramic) detector arrays positioned at a 90-degree offset, acquiring comprehensive volumetric datasets at twice the speed of single-source systems.
-
83 ms Temporal Resolution: Freezes fast-moving coronary arteries completely. Multisector reconstruction algorithms can further enhance temporal clarity down to 42 ms.
-
Simultaneous Dual-Energy Scanning: Acquires low- and high-kV data sets simultaneously from the same spatial location and phase, enabling precise virtual non-contrast images, gout differentiation, and kidney stone characterization.
-
Twin STRATON™ X-Ray Tubes: Heavy-duty, direct-anode liquid-cooled tubes engineered with zero-heat-storage design principles, supplying continuous high-mA output during demanding multi-phase acquisitions.
-
160 kW High-Power Generator Reserve: Combines dual 80 kW generators to guarantee stable, robust photon output across large bariatric patients and dense anatomical regions.
-
Extended 78 cm Gantry Bore: Provides wider physical clearance to accommodate claustrophobic patients, trauma frames, and interventional positioning devices comfortably.
Technical Specifications
| Parameter | Specification |
| Gantry Architecture | Dual-Source (2 x STRATON™ Tubes / 2 x 32-Row Detectors) |
| Slice Reconstructive Capability | 64-Slice Equivalent Volume per Rotation (via z-Sharp flying focal spot) |
| Native Temporal Resolution | 83 ms in-plane (down to 42 ms with multisector processing) |
| Gantry Aperture (Bore) | 78 cm Clearance |
| Gantry Rotation Speed | Down to 0.33 seconds full 360° rotation |
| Total Generator Power | 160 kW (2 x 80 kW independent generators) |
| Spatial Resolution | Up to 0.33 mm isotropic high-contrast resolution |
| Scan Range Capacity | Up to 200 cm continuous z-axis range |
| Patient Table Load | Supports up to 200 kg standard operational weight |



Reviews